Failure of cemented hip implants under complex loading : experimental and numerical analysis

Loading...
Thumbnail Image

Date

Journal Title

Journal ISSN

Volume Title

Publisher

Trinity College (Dublin, Ireland). Department of Mechanical and Manufacturing Engineering

Access

openAccess

Embargo end date

Citation

John Robert Britton, 'Failure of cemented hip implants under complex loading : experimental and numerical analysis', [thesis], Trinity College (Dublin, Ireland). Department of Mechanical and Manufacturing Engineering, 2004, pp 196

Abstract

A new generation of experimental pre-clinical testing methods for cemented hip replacement femoral components have been proposed, and preliminarily validated by comparing a 'good' prosthesis design with a 'bad' prosthesis design. The basis for these new tests is the measurement of relative motion between im plant and bone over time while loading is applied. However, it is yet to be proven whether these next generation tests can satisfactorily distinguish between the performance of several prosthesis designs. Additionally, opinions vary widely as to the necessary loading conditions required, particularly over whether or not muscle loading should be simulated.

Description

Endorsement

Review

Supplemented By

Referenced By

Qualification name: Doctor of Philosophy (Ph.D.)
Publisher: Trinity College (Dublin, Ireland). Department of Mechanical and Manufacturing Engineering
Type of material: thesis