Cost & Capability Compromises in STEM Instrumentation for Low-Voltage Imaging
Loading...
Files
Date
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Access
openAccess
Embargo end date
Citation
Quigley, Frances and McBean, Patrick and O'Donovan, Peter and Jones, Lewys, Cost & Capability Compromises in STEM Instrumentation for Low-Voltage Imaging, Microscopy and Microanalysis, 2022, 1437 - 1443
Abstract
When characterising beam-sensitive materials in the scanning transmission electron microscope (STEM), low-
dose techniques are essential for the reliable observation of samples in their true state. A simple route to
minimise both the total electron-dose and the dose-rate is to reduce the electron beam-current and/or raster the
probe at higher speeds. At the limit of these settings, and with current detectors, the resulting images suffer
from unacceptable artefacts including; signal-streaking, detector-afterglow, and poor signal-to-noise ratios
(SNR). In this manuscript we present an alternative approach to capture dark-field STEM images by pulse-
counting individual electrons as they are scattered to the annular dark-field (ADF) detector. Digital images
formed in this way are immune from analogue artefacts of streaking or afterglow and allow clean, high-SNR
images to be obtained even at low beam-currents. We present results from both a ThermoFisher FEI Titan G2
operated at 300kV and a Nion UltraSTEM200 operated at 200kV, and compare the images to conventional
analogue recordings. ADF data are compared with analogue counterparts for each instrument, a digital
detector-response scan is performed on the Titan, and the overall rastering efficiency is evaluated for various