Predictions of the Compressible Fluid Model and its Comparison to Experimental Measurements of Q Factors and Flexural Resonance Frequencies for Microcantilevers
Loading...
Files
Date
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Access
openAccess
Embargo end date
Citation
Jensen, J. and Hegner, M., Predictions of the Compressible Fluid Model and its Comparison to Experimental Measurements of Q Factors and Flexural Resonance Frequencies for Microcantilevers, Journal of Sensors, 2012, 2012, 258381-
Abstract
The qualitative agreement between experimental measurements of the 𝑄 factors and flexural resonance frequencies in air of microcantilevers and calculations based on the compressible fluid model of Van Eysden and Sader (2009) is presented. The 𝑄 factors and resonance frequencies observed on two sets of cantilever arrays were slightly lower than those predicted by the model. This is attributed to the individual design and geometry of the microfabricated hinged end of the cantilever beams in the array.
Description
PUBLISHED
Inverse behaviour of Quality factors observed in higher modes in the gas phase
Inverse behaviour of Quality factors observed in higher modes in the gas phase
Collections
Endorsement
Review
Supplemented By
Referenced By
Keywords
Sponsor: Science Foundation Ireland (SFI)
Grant Number: SFI08/CE/I1432
Sponsor: Science Foundation Ireland (SFI)
Grant Number: SFI/09IN/1B2623
Author's Homepage: http://people.tcd.ie/hegnerm
Type of material: Journal Article