Bilateral 23Na MR imaging of the breast and quantification of sodium concentration
File Type:
PDFItem Type:
Journal ArticleDate:
2014Author:
Access:
openAccessCitation:
Danisch, Kalayciyan, Wetterling, Schad, Bilateral 23Na MR imaging of the breast and quantification of sodium concentration, ZEITUNG FUER MEDIZINISCHE PHYSIK, 24, 1, 2014, 65-72Download Item:
Abstract:
A novel setup for 23 Na MRI, which allows bilateral imaging
of the breast, is presented. For this purpose a figure-
eight receive-only 23 Na surface coil was developed. For
our experiments on three samples with NaCl solutions of
different sodium concentrations and two female subjects
we used an asymmetric birdcage coil in transmit mode
and the developed surface coil for receiving the signal
at 3 T. Imaging of the samples showed the applicabil-
ity of the employed normalization method for measuring
the distribution of sodium concentration. In a sample
of concentration [Na + ] = 51 mM we achieved SNR = 70
at a nominal isotropic resolution of 2,5 mm (T R =66 ms,
TE =0,6 ms, TA = 20 min). Furthermore we showed that by
means of this setup it is possible to quantify the sodium concentration in breast tissue (TSC) of a female subject with an
accuracy of 23% (T R =150 ms, T E =0,5 ms, T A =45 min).
Author's Homepage:
http://people.tcd.ie/wetterfDescription:
PUBLISHED
Author: Wetterling, Friedrich
Other Titles:
Bilaterale 23Na-MR-Bildgebung der Mamma und Quantifizierung der NatriumkonzentrationBilateral 23Na MR imaging of the breast and quantification of sodium concentrationType of material:
Journal ArticleSeries/Report no:
ZEITUNG FUER MEDIZINISCHE PHYSIK;24;
1;
Availability:
Full text availableSubject (TCD):
Cancer , Next Generation Medical DevicesDOI:
http://dx.doi.org/10.1016/j.zemedi.2013.07.004Metadata
Show full item recordLicences: