Title: Development of a vessel-mimicking material for use in anatomically realistic Doppler flow phantoms
Author: FAGAN, ANDREW JOHN
Sponsor: 
Name Grant Number
PG/07/107/23895

Author's Homepage: http://people.tcd.ie/faganan
Keywords: Polyvinyl alcohol cryogel
anatomically realistic flow phantom
vessel mimic
acoustic and mechanical characterisation
Issue Date: 2011
Citation: D.M. King, C.M. Moran, J. McNamara, A.J. Fagan and J.E. Browne, Development of a vessel-mimicking material for use in anatomically realistic Doppler flow phantoms, Ultrasound in Medicine and Biology, 37, 5, 2011, 813 - 826
Series/Report no.: Ultrasound in Medicine and Biology
37
5
Abstract: Polyvinyl alcohol cryogel, (PVA-C) is presented as a vessel mimicking material for use in anatomically realistic Doppler flow phantoms. Three different batches of 10 % wt PVA-C containing (i) PVA-C alone, (ii) PVA C with anti-bacterial agent and (iii) PVA-C with silicon carbide particles were produced, each with 1 to 6 freeze-thaw cycles. The resulting PVA-C samples were characterized acoustically (over a range 2.65 – 10.5 MHz) and mechanically in order to determine the optimum mixture and preparation for mimicking the properties of healthy and diseased arteries found in vivo. This optimum mix was reached with the PVA-C with anti-bacterial agent sample, prepared after 2 freeze/thaw cycles, which achieved a speed of sound of 1538 ± 5 m s 1 and a Young’s elastic modulus of 79 ± 11kPa. This material was used to make a range of anatomically-realistic flow phantoms with varying degrees of stenoses, and subsequent flow experiments revealed that higher degrees of stenoses and higher velocities could be achieved without phantom rupturing compared to a phantom containing conventional wall-less vessels.
Description: PUBLISHED
URI: http://hdl.handle.net/2262/56863
Related links: http://dx.doi.org/10.1016/j.ultrasmedbio.2011.02.012
Appears in Collections:Physics (Scholarly Publications)

Files in This Item:

File Description SizeFormat
Final Version.docPublished (author's copy) - Peer Reviewed241 kBMicrosoft WordView/Open


This item is protected by original copyright


Please note: There is a known bug in some browsers that causes an error when a user tries to view large pdf file within the browser window. If you receive the message "The file is damaged and could not be repaired", please try one of the solutions linked below based on the browser you are using.

Items in TARA are protected by copyright, with all rights reserved, unless otherwise indicated.

 

Valid XHTML 1.0! DSpace Software Copyright © 2002-2010  Duraspace - Feedback