Assessment of the accuracy of an ultrasound elastography liver scanning system using a PVA-cryogel phantom with optimal acoustic and mechanical properties
S. Cournane, L.M. Cannon, J.E. Browne, A.J. Fagan, Assessment of the accuracy of an ultrasound elastography liver scanning system using a PVA-cryogel phantom with optimal acoustic and mechanical properties, Physics in Medicine and Biology, 55, 19, 2010, 5965-5983
Series/Report no.:
Physics in Medicine and Biology 55 19
Abstract:
The accuracy of a transient elastography liver-scanning ultrasound system was assessed using a
novel application of PVA-cryogel as a tissue-mimicking material with acoustic and shear
elasticity properties optimized to best represent those of liver tissue. Although the
liver-scanning system has been shown to offer a safer alternative for diagnosing liver cirrhosis
through stiffness measurement, as compared to the liver needle biopsy exam, the scanner’s
accuracy has not been fully established. The Young’s elastic modulus values of 5-6wt%
PVA-cryogel phantoms, also containing glycerol and 0.3μm Al2O3 and 3μm Al2O3, were
measured using a ‘gold standard’ mechanical testing technique and transient elastography. The
mechanically measured values and acoustic velocities of the phantoms ranged between
1.6-16.1kPa and 1540-1570m/s, respectively, mimicking those observed in liver tissue. The
values reported by the transient elastography system overestimated the Young’s elastic modulus
values representative of the progressive stages of liver fibrosis by up to 32%. These results were
attributed to the relative rather than absolute nature of the measurement arising from the singlepoint
acoustic velocity calibration of the system, rendering the measurements critically
dependent on the speed of sound of the sample under investigation. Given the wide range of
acoustic velocities which exist in the liver, spanning healthy tissue to cirrhotic pathology,
coupled with the system’s assumption that the liver is approximately elastic when it is rather
highly viscoelastic, care should be exercised when interpreting the results from this system in
patient groups.
Please note: There is a known bug in some browsers that causes an
error when a user tries to view large pdf file within the browser window.
If you receive the message "The file is damaged and could not be
repaired", please try one of the solutions linked below based on the
browser you are using.
Items in TARA are protected by copyright, with all rights reserved, unless otherwise indicated.